Development of a near-infrared spectroscopy interface able to assess oxygen recovery kinetics in the right and left sides of the pelvic floor

dc.contributor.authorMacnab, Andrewen_ZA
dc.contributor.authorStothers, Lynnen_ZA
dc.contributor.authorDeegan, Emilyen_ZA
dc.date.accessioned2021-10-19T10:16:24Z
dc.date.available2021-10-19T10:16:24Z
dc.date.issued2019
dc.descriptionCITATION: Macnab, A., Stothers, L. & Deegan, E. 2019. Development of a near-infrared spectroscopy interface able to assess oxygen recovery kinetics in the right and left sides of the pelvic floor. Journal of Biomedical Optics, 24(7):075003, doi:10.1117/1.JBO.24.7.075003.
dc.descriptionThe original publication is available at https://www.spiedigitallibrary.org/journals/journal-of-biomedical-optics
dc.description.abstractNear-infrared spectroscopy (NIRS) muscle oxygenation data are relied on in sports medicine. Many women with urinary incontinence (UI) have dysfunctional pelvic floor muscles (PFMs) but their evaluation lacks such measures; a transvaginal NIRS interface would enable the PFM to be interrogated. Paired miniature fiber-optic cables were configured on a rigid foam insert so their emitter detector arrays with an interoptode distance of 20 mm apposed the right and left inner sides of a disposable clear plastic vaginal speculum, and linked to a standard commercial NIRS instrument. Measurement capability was assessed through conduct of three maximum voluntary contractions (MVCs) and one sustained maximum voluntary contraction of the PFM with calculation of HbDiff (½RT), a validated muscle reoxygenation kinetic parameter. In all four asymptomatic controls, mean age 40, mean BMI 21.4, MVCs were associated with changes in PFM oxyhemoglobin (O2Hb), deoxyhemoglobin (HHb) concentration, and their difference (HbDiff) comparable to those in voluntary muscle sports medicine studies. NIRS data during recovery (reoxygenation) allowed calculation of HbDiff (½RT). New techniques are called for to evaluate UI. This NIRS interface warrants further development as the provision of quantitative reoxygenation kinetics offers more comprehensive evaluation of patients with PFM dysfunction.en_ZA
dc.description.urihttps://www.spiedigitallibrary.org/journals/journal-of-biomedical-optics/volume-24/issue-07/075003/Development-of-a-near-infrared-spectroscopy-interface-able-to-assess/10.1117/1.JBO.24.7.075003.full
dc.description.versionPublisher's version
dc.format.extent6 pages
dc.identifier.citationMacnab, A., Stothers, L. & Deegan, E. 2019. Development of a near-infrared spectroscopy interface able to assess oxygen recovery kinetics in the right and left sides of the pelvic floor. Journal of Biomedical Optics, 24(7):075003, doi:10.1117/1.JBO.24.7.075003
dc.identifier.otherdoi:10.1117/1.JBO.24.7.075003
dc.identifier.urihttp://hdl.handle.net/10019.1/123252
dc.language.isoen_ZAen_ZA
dc.publisherSociety of Photo-optical Instrumentation Engineers
dc.rights.holderAuthors retain copyright
dc.subjectNear-infrared spectroscopyen_ZA
dc.subjectPelvis floor -- Musclesen_ZA
dc.subjectOxygen in the body -- Measurementen_ZA
dc.subjectUrinary incontinence -- Measurement -- Equipment and suppliesen_ZA
dc.titleDevelopment of a near-infrared spectroscopy interface able to assess oxygen recovery kinetics in the right and left sides of the pelvic flooren_ZA
dc.typeArticleen_ZA
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
macnab_development_2019.pdf
Size:
967.63 KB
Format:
Adobe Portable Document Format
Description:
Download article
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: