'n Vergelykende ondersoek tussen twee kapasitorontlandings-x-straaleenhede
Date
1980-11
Authors
Frost, J. M.
Collins, P. D.
Van der Merwe, E. J.
Journal Title
Journal ISSN
Volume Title
Publisher
Health and Medical Publishing Group (HMPG)
Abstract
Two capacitor discharge X-ray units of different make were compared. The same radiological examinations were carried out with both units and the radiation exposure for every examination was measured on a Rando phantom with calibrated LiF-TLD Teflon discs. A large difference in radiation exposure for the same image quality was found between the two units. Factors such as the use of rare earth intensifying screens with capacitor discharge X-ray units, independent tube current settings and automatic charge replenishment are stressed.
Description
The original publication is available at http://www.samj.org.za
Keywords
Diagnostic imaging -- Equipment and supplies, X-rays -- Equipment and supplies, Capacitor discharge X-ray units